Recall Z-0518-2020 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Becton Dickinson & Company

Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 S

DateNov 4, 2019
StatusTerminated
ReasonThe indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only
Root causeUnder Investigation by firm
ActionBD issued to US consignees via FedEx beginning November 4, 2019 Urgent Medical Device Notification, stating reason for recall , health risk and action to take: Until the issue is resolved, it is recommended that all Site~Rite 8 Ultrasound Systems be operated using AC power at all times during use in a clinical procedure. If use of AC power is not practical for your clinical setting, we recommend the use of an external re-chargeable battery. The listing of batteries have been qualified for use with the Site~Rite 8 Ultrasound System and are available for purchase through common retailers were i
Quantity2215 units
DistributionNationwide Foreign: UK, Belgium, Australia, Japan, Canada, Taiwan, Colombia, Chile, Philippines, Thailand, Brazil, Hong Kong, Chile, Argentina, Mexico, India, Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152554Site-Rite 8 Ultrasound System, Site-Rite 8 Ultrasound System with Pinpoint GT Technology
C.R. Bard, Inc., Dec 14, 2015