Recall Z-0608-2020 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips North America, LLC

HeartStart XL+ Defibrillator/Monitor, Model 861290

DateOct 3, 2019
StatusTerminated
ReasonDevice may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
Root causeProcess control
ActionPhilips Healthcare issued Medical Device Correction Notices dated 10/3/2019 to customers via certified mail. Customers are advised to take the following actions: - The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this notice. - If you identify a device that exhibit these behaviors, please remove the device from service and contact Philips to request device. - Fill out, sign and send the reply form that is provided as the last page of the letter.
Quantity24,738 units
DistributionUS Nationwide