Recall Z-0610-2020 ยท FRN Pump, Infusion

Smiths Medical ASD Inc.

Medfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. The pump offers a variety of delivery modes programmable to meet specific patient care needs.

DateDec 12, 2019 (initiated Oct 28, 2019)
StatusTerminated
ReasonThere is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.
Root causeSoftware design
ActionAn Urgent Medical Device recall letter dated 10/28/2019 was sent to customers via email beginning 10/28/2019. The letter identified affected product, stated the issue, and provided instructions for customers to return the pumps. A response form was asked to be returned. Questions can be directed to fieldactions@smiths-medical.com.
Quantity626 devices
DistributionCO, MD, NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111386MEDFUSION MODEL 4000 SYRINGE INFUSION PUMP, PHARMGUARD TOOLBOX 2 MEDICATION SAFETY SOFTWARE AND SUPPORTED SYRINGES (PSS)
Smiths Medical MD, Inc., Aug 29, 2011