Recall Z-0686-2020 ยท LCP Assay, Glycosylated Hemoglobin

Polymer Technology Systems

CVS Health At Home A1C Test Kit, Software Version Revision D, REF Number 3059

DateDec 2, 2016
StatusTerminated
ReasonA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Root causeSoftware design
ActionA Customer Bulletin dated November 2016 was distributed to customers.
Quantity8285
DistributionThe products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090413A1CNOW+ (10 TEST KIT, PROFESSIONAL USE) MODEL 3024, A1CNOW+(20 TEST KIT, PROFESSIONAL USE) MODEL 3021, A1CNOW SELF CHEC
Bayer Healthcare, LLC, May 14, 2009