ClearanceBrief
Recall Z-0683-2020 ยท LCP Assay, Glycosylated Hemoglobin

Polymer Technology Systems

A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062

DateDec 2, 2016
StatusTerminated
ReasonA numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Root causeSoftware design
ActionA Customer Bulletin dated November 2016 was distributed to customers.
Quantity500
DistributionThe products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090413A1CNOW+ (10 TEST KIT, PROFESSIONAL USE) MODEL 3024, A1CNOW+(20 TEST KIT, PROFESSIONAL USE) MODEL 3021, A1CNOW SELF CHEC
Bayer Healthcare, LLC, May 14, 2009