Recall Z-0603-2020 ยท HRS Plate, Fixation, Bone

Miami Device Solutions, LLC.

Proximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps intended to provide internal fixation of fractures of the proximal humerus. The polyaxial screws have a range of motion of +/- 20. The screws can be simply lagged or locked to the plate with the addition of a locking cap. The plates are side speci

DateJan 15, 2019
StatusTerminated
ReasonRight Proximal Humerus Plates. Product component parts MDS130108R, may be improperly identified as left.
Root causeProcess control
ActionThe firm notified their consignees by email beginning on 01/15/2019 requesting the examination of the Proximal Humerus Plates for mismarked units. Return of the units is ongoing.
Quantity15 units
DistributionUS Nationwide distribution in the states of FL, RI, MA, PA, NY, OH, GA, WA, AL, CA, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141493PROXIMAL HUMERUS PLATING SYSTEM
Miami Device Solutions, Mar 20, 2015