Recall Z-0728-2020 ยท NIO Stent, Iliac

Cardinal Health Inc.

S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diamet

DateNov 14, 2019
StatusOpen, Classified
ReasonCordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Root causeUnder Investigation by firm
ActionCustomers were notified with a Medical Device Correction letter. The letter was sent via overnight delivery beginning 11/14/2019. The letter identified affected product, provided a description of the problem, and asked that a copy of the letter be retained where the product is stored. Customer were asked to completed and return an Acknowledgement Form. Questions can be directed to the Cordis Endovascular Medical Information Line 877-DEVICE5 (877-338-4235) Monday through Friday 9:00 am - 5:00 pm EST.
Quantity267,247 units total (262,641 OUS, 4,606 US)
DistributionWorldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecua
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.