Recall Z-0612-2020 ยท HWA Impactor

Enztec Limited

REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

DateJan 24, 2020 (initiated Aug 16, 2019)
StatusTerminated
ReasonThe femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.
Root causeUnknown/Undetermined by firm
ActionOn August 16, 2019, a Urgent Medical device recall for Product Correction" letter was emailed to affected consignees. In addition to informing the consignees about the recall, the
Quantity312 Femoral Impactors