ClearanceBrief
Recall Z-0590-2020 ยท KRA Catheter, Continuous Flush

Cook Inc.

CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures.

DateDec 9, 2019 (initiated Nov 8, 2019)
StatusTerminated
ReasonA manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
Root causeProcess control
ActionUrgent Medical Device Recall notification letters dated 11/8/19 were sent. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Ac
Quantity58
DistributionDistribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161801CrossCath Support Catheter
Cook Incorporated, Jul 29, 2016