Recall Z-0761-2020 ยท MMI Immunoassay Method, Troponin Subunit

Abbott Point Of Care Inc.

i-STAT cTnI cartridge

DateOct 16, 2019
StatusTerminated
Reasoni-STAT cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
Root causeDevice Design
ActionOn October 16, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail. It explains Abbott Point of Care has made the decision to transition U.S. customers using the i-STAT cTnI blue cartridge, list number 06P23-25, to the i-STAT cTnI white cartridge, list number 03P90-25, beginning in November 2019. The i-STAT cTnI blue cartridge, list number 06P23-25, will be discontinued for sale in the United States as of January 1, 2020. It further explains to customers while the i-STAT cTnI blue cartridges can continue to be used for patient testing, APOC recommends that a
Quantity3,465,875
DistributionUS: AK AL AZ CA CT DC FL GA HI IL IN KS KY MD MI MS NC NE NH NM NV NY OH OK OR PA SC SD TN TX VA WA WV WY