Recall Z-0578-2020 ยท DQY Catheter, Percutaneous

Boston Scientific Corporation

Guidezilla II Guide Extension Catheter Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493933515060, 6F; 2. REF/Catalog No H7493933515070, 7F.

DateOct 24, 2019
StatusTerminated
ReasonCertain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
Root causeProcess control
ActionThe firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: BSCFieldActionCenter@bsci.com OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases
Quantity14,145 devices total
DistributionNationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163314Guidezilla II Guide Extension Catheter; Guidezilla II LONG Guide Extension Catheter
Boston Scientific Corporation, Mar 23, 2017