Recall Z-0466-2020 ยท DQO Catheter, Intravascular, Diagnostic

Cook Inc.

Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.

DateOct 21, 2019
StatusTerminated
ReasonThe potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
Root causeProcess change control
ActionOn October 21, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail instructing them to: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Ev
Quantity25
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133130BEACON TIP TORCON NB ADVANTAGE CATHETER, TORCON NB ADVANTAGE CATHETER
Cook, Inc., Jan 9, 2015