Recall Z-0498-2020 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips North America, LLC

HeartStart XL+ Defibrillator/Monitor, Model 861290

DateNov 19, 2019 (initiated Oct 16, 2019)
StatusTerminated
ReasonThe rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from the setting selected by the user.
Root causeProcess control
ActionOn October 16, 2019, Philips Healthcare issued Urgent Medical Device Correction notices to U.S. customers via certified mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Actions to be taken by the customer: 1) The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this Notice. 2) Continue to perform Shift Checks and Operational checks as recommended in the Instructions for Use (IFU) as this reduces the risk of a failure during use. 3) If you identify a device that exhibit any of t
Quantity23,749 Units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110825PHILIPS HEARTSTART XL+, PHILIPS HEARTSTART EXTERNAL PADDLES, PHILIPS MULTIFUNCTION ELECTRODES
Philips Medical Systems, Feb 16, 2012