Recall Z-0771-2020 ยท NQE Microscope, Specular

Nidek, Inc.

NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.

DateFeb 25, 2019
StatusTerminated
ReasonThe electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
Root causeDevice Design
ActionThe recalling firm issued letters dated 2/7/2019 via FedEx explaining the issue, informing the consignee there was no risk associated with use of the device, and that the firm is preparing for the replacement parts in order to schedule for the correction to be made.
Quantity13 devices
DistributionUS Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130565SPECULAR MICROSCOPE CEM-530
Nidek Co., Ltd., Nov 27, 2013
K151706Specular Microscope CEM- 530
Nidek Co., Ltd., Mar 17, 2016
K173980Specular Microscope CEM-530
Nidek Co., Ltd., Mar 14, 2018