ClearanceBrief
Recall Z-0540-2020 ยท FMT Warmer, Infant Radiant

GE Healthcare, LLC

Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer

DateDec 3, 2019 (initiated Oct 11, 2019)
StatusTerminated
ReasonGiraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Root causeDevice Design
ActionThe firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an
Quantity3467 units
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152809Giraffe Incubator Carestation CS1
Ohmeda Medical, A Division of Datex-Ohmeda, Inc., Dec 31, 2015