Recall Z-0511-2020 ยท PLV Dental Stereotaxic Instrument

Neocis Inc.

Neocis Guidance System Fiducial Array - Product Usage: The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intraoperative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical

DateOct 4, 2019
StatusTerminated
ReasonThe Fiducial Array may mismatch their measurement files resulting in failed landmark check.
Root causeProcess control
ActionThis firm initiated the recall by visit, field service personnel to each affected to customer to retrieve the nonconforming parts beginning on 10/04/2019. The firm followed this action officially notified customers, via letter on sent on October 15, 2019.
Quantity3 units
DistributionUS Nationwide distribution in the states of AR, OK, NY.
FDA recordOpen on fda.gov