ClearanceBrief
Recall Z-0791-2020 ยท NDP Accessories, Implant, Dental, Endosseous

Implant Direct Sybron Manufacturing, LLC

ImplantDirect InterActive Precision IO Scan Adapter, Part #6530-09PT, Platform: 3.0mm, Non-Sterile, Rx. The firm name on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA.

DateNov 28, 2018
StatusCompleted
ReasonThe peek part was loose on the titanium base of the device which could result in it falling into a patient's mouth during a procedure resulting in aspiration or ingestion.
Root causeProcess control
ActionThe firm first issued recall letters on 11/28/2018 via FedEx for Part number 6534-09PT and then on 12/28/2018, issued recall letters for Part number 6530-09PT via FedEx. On 2/14/2019, the firm issued another letter to their consignees that added a risk statement regarding the recall of both part numbers (6534-09PT and 6530-09PT).
Quantity232 devices
DistributionDistribution was made to AL, AZ, CA, FL, GA, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, and VA. There was no government/military distribution. Foreign distribution was made to Canada, Armenia, Australia, Denmark, France, Germany, Great Britain, Iceland, Italy,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.