ClearanceBrief
Recall Z-0689-2020 ยท FMI Needle, Hypodermic, Single Lumen

Teleflex Medical

Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, o

DateOct 8, 2019
StatusOpen, Classified
ReasonThe safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
Root causePackage design/selection
ActionUrgent Medical Device Correction notification letters dated 10/8/19 were sent to customers. Customer Action: Our records indicate you have received product that is subject to this correction. Place a copy of this notice with the product to ensure all users are aware of the need to perform this inspection. Removing the kits from the shelf box and placing them into bags (e.g. backpacks) could increase the potential for caps to become loose and product should be visually inspected again at point of use. Our records indicate that you have received products that are subject to this action. We ar
Quantity526,000 total
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141117EZ-IO INTRAOSSEOUS INFUSION SYSTEM
Vidacare, LLC, Jul 8, 2014