ClearanceBrief
Recall Z-0414-2020 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Randox Laboratories, Limited

Randox Liquid Cardiac Controls Catalogue Numbers CQ5053.

DateOct 4, 2019
StatusTerminated
ReasonRemoval of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
Root causeUnder Investigation by firm
ActionOn October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy
Quantity6,702
DistributionUS Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110534RANDOX LIQUID CARDIAC CONTROL LEVELS 1, 2 AND 3
Randox Laboratories Limited, Jan 23, 2012