Recall Z-0759-2020 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

GE Healthcare, LLC

CARESCAPE Respiratory Module E-sCAiOVE

DateJan 10, 2020 (initiated Sep 27, 2019)
StatusTerminated
ReasonDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Root causeProcess control
ActionThe firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configu
Quantity14 unit
DistributionUS OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia,