Recall Z-0679-2020 ยท LPS Urinary Homocystine (Nonquantitative) Test System

Roche Diagnostics Operations, Inc.

Roche Homocysteine-In vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems Catalog Number: 06542921190 - - Product Usage: The assay can assist in the diagnosis of patients suspected of having hyperhomocysteinemia or homocystinuria.

DateOct 4, 2019
StatusTerminated
ReasonHomocysteine Reagent, Calibration Failures and Quality Control Recovery Issues on cobas c 701 and 702 modules
Root causeDevice Design
ActionThe Urgent Medical Device Correction (UMDC) was sent via Roche issued Urgent Medical Device notification via FedEx (signature required) to the consignee list on 10/04/2019. The letter states issue, health risk nd action to take: Discontinue use and discard the remaining affected product in inventory according to local waste guidelines. Because a replacement lot is currently not available, Roche will issue credit for Homocysteine cobas c packs, lot number 422313, remaining in your inventory. If customers have another validated system (i.e. cobas c501/502) in their lab and wish to set up the
Quantity82 units (US)
DistributionUS Nationwide distribution in the states of NJ, HI, IA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113793HOMOCYSTEINE TEST
Roche, Jun 5, 2012