ClearanceBrief
Recall Z-0216-2020 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

DateNov 1, 2019 (initiated Sep 19, 2019)
StatusTerminated
ReasonAn issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Root causeNonconforming Material/Component
ActionOn September 19, 2019 the firm distributed Urgent Field Safety Notice (FSN) Medical Device Correction letters by direct communication from Philips Field Service Engineers instructing them to: - Discontinue system use until further notice - Inform those who need to be aware within your organization or any organization where the potentially affected devices have been transferred (If appropriate). - Maintain this notice with your system Instructions for Use (IFU) until the correction is made to the system. ACTIONS PLANNED BY PHILIPS: Philips Healthcare is distributing this FSN to all affected cu
Quantity1167 in total
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982911FORTE
Adac Laboratories, Nov 16, 1998