Recall Z-0270-2020 ยท LXH Orthopedic Manual Surgical Instrument

Pega Medical Inc.

SCFE (slipped capital femoral epiphysis) Knobs for Driver, 6.5mm and 7.3 mm Orthopedic surgical instrument component, for hip surgery

DateOct 2, 2019
StatusTerminated
ReasonThere is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
Root causeDevice Design
ActionPega Medical notified customers of the recall on about 10/02/2019, via "Urgent Field Safety Notice - Recall" letter. Customers were informed that SCFE Drivers, 6.5mm and 7.3 mm of the Free Gliding SCFE Screw System had issues with the knobs. Specifically the knobs have the potential to have loosened screws, or screws that have fallen out, as a result of the ultrasonic cleaning of the knobs. Customers were instructed to identify affected devices in inventory and return them to Pega Medical as soon as possible to address the issue. There returned devices can be returned via the FedEx infor
Quantity31 devices
DistributionNationwide distribution to CA, FL, GA, HI, LA, MA, MO, NC, NY, OK. International distribution to AUSTRALIA, CANADA, DENMARK, GERMANY, ISRAEL, MEXICO, ROMANIA, RUSSIA, UNITED ARAB EMIRATES, UNITED KINGDOM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.