ClearanceBrief
Recall Z-0821-2020 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Isopure Corp

Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

DateJan 19, 2020 (initiated Apr 19, 2019)
StatusTerminated
ReasonThere is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Root causeNonconforming Material/Component
ActionThe only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.
Quantity15
DistributionThe products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120005ACDS
Isopure, Corp., Mar 29, 2012