ClearanceBrief
Recall Z-0235-2020 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Mindray DS USA, Inc. dba Mindray North America

Accutorr 3 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical phys

DateOct 1, 2019
StatusTerminated
ReasonNIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
Root causeNo Marketing Application
ActionMindray communicated by letter on 10/1/19 to consignees and distributors by certified postal mail with return receipt verification: to consignees that have affected vital sign monitor; communication to consignees that purchased directly NIBP valve assembly boards; communication to distributors that have affected vital sign monitor. Consignees were instructed to contact Mindrays Repair Center to arrange for affected units to be returned for a software update.
Quantity13 units
DistributionUS Nationwide distributions.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132037VITAL SIGNS MONITOR
Mindray North America, May 16, 2014