Recall Z-0176-2020 ยท LMA Digitizer, Image, Radiological
Brainlab AG
Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DATA MANAGER 2.2.0 LICENSE ONLY 25100-11 - PATIENT DATA MANAGER 2.2.1 LICENSE ONLY 25100-12 - PATIENT DATA MANAGER 2.3 INSTALLER 25100-12A - PATIENT DATA MANAGER 2.3.1 INSTALLER 25100-13 - PATIENT DATA MANAGER 2.3 LICENSE 25100-12B - PATIENT DATA MANAGER 2.4.0 INSTALLER 2510
| Date | Sep 25, 2019 |
| Status | Terminated |
| Reason | The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances. |
| Root cause | Software design |
| Action | As of September 25, 2019 for new installations or updates/upgrades to the affected SW version, the Field Safety Notice / Product Notification letter is hand delivered to the customer during installation (before clinical use) by the according Brainlab Service representative All other consignees were notified by email on 10/04/2019. The letter details how the even occurs and provides instruction on how to avoid the issue until a software revision is disseminated. |
| Quantity | 1230 units |
| Distribution | worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.