Recall Z-0338-2020 ยท HTN Washer, Bolt Nut
Zimmer Biomet, Inc.
cable button for NCB Polyaxial Locking Plate, 2.5 mm Hex Drive, Item No. 47223206001
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5459583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K120772 | NCB PERIPROSTHETIC TROCHANTER PLATES AND SCREWS, NCB CABLE BUTTON FOR NCB OPLYAXIAL LOCKING PLATE Zimmer GmbH, Jun 12, 2012 |