Recall Z-0334-2020 ยท KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Zimmer Biomet, Inc.
Metasul Taper Liner, various sizes, Item Nos. 00877000528 00877000628 00877000732 00877000832 00877000936 00877001036 00877001140 00877001240 00877001340 00877001440 00877001540 00877001640 00877001740 00877001840 00877001940 00877002040 00877002140 00877002240 00877002340
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K091973 | METASUL TAPER LINERS, METASUL FEMORAL HEADS Zimmer GmbH, Nov 17, 2009 |