Recall Z-0329-2020 ยท HRS Plate, Fixation, Bone

Zimmer Biomet, Inc.

Various temporary devices used during the implantation of the device system: bone fixation plates and screws, drill bits, manual surgical instruments for general use. Item Nos. 0202024301 0202024325 0202024330 0202024342 0202024344 0202024346 0202024360 0202024362 47225806700

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100111NCB PERIPROSTHETIC FEMUR POLYAXIAL LOCKING PLATE SYSTEM
Zimmer, Inc., Apr 12, 2010