Recall Z-0327-2020 · KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Zimmer Biomet, Inc.

CABLE-READY SHORT GTR and CABLE-READY LONG GTR, Item Nos. 00223200201 00223200202

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151907Cable-Ready® Cable Grip System: Cable-Ready® GTR System, Cable-Ready® 1.8mm Cerclage Cables
Zimmer, Inc., Dec 30, 2015