Recall Z-0325-2020 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer Biomet, Inc.
Various Knee Systems, Item Nos. 32855411838 32855411840 32855411842 32855411843 32855420307 32855420308 32855420309 32855420310 32855420315 32855430491 32855430497 32855430499 32855430500 32855430502 32855430503 32855430511 32855430526 32855430530 32855430531 32855430536 32855430537 32855430538 32855430539 32855430541 32855430543 32855430546 32855430548 32855430549 32855430550 32855430551 32855430
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall