Recall Z-0319-2020 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Zimmer Biomet, Inc.
CLS Brevius Stem Kinectiv, hip prosthesis, various sizes. Item Nos. 0100296050 0100296060 0100296070 0100296080 0100296090 0100296100 0100296112 0100296125 0100296137 0100296150 0100296162 0100296175 0100296200
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |