Recall Z-0317-2020 ยท HRS Plate, Fixation, Bone

Zimmer Biomet, Inc.

Periarticular locking plate elbow-sterile, Item Nos. 47235800503 47235800509 47235800511 47235800515 47235800519 47235800603 47235800605 47235800607 47235800609 47235800611 47235800615 47235800619 47235800705 47235800711 47235800715 47235800803 47235800805 47235800807 47235800811 47235800815 47235800911 47235800913 47235800915 47235801209 47235801211 47235801213 47235801215 47235802305 47235802307

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082078ZIMMER PERIARTICULAR LOCKING PLATE SYSTEM: DISTAL HUMERAL AND PROXIMAL ULNA PLATES
Zimmer, Inc., Nov 3, 2008
K141734ZIMMER PERIARTICULAR LOCKING SYSTEM
Zimmer, Inc., Aug 13, 2014