ClearanceBrief
Recall Z-0315-2020 ยท HRS Plate, Fixation, Bone

Zimmer Biomet, Inc.

PERI DIST VOLAR/DORSAL T/DORSAL DELTA RADIAL, Item Nos. 00234801303 00234801304 00234801306 00234801308 00234801310 00234801312 00234801403 00234801404 00234801406 00234801408 00234801410 00234801412 00234802102 00234802104 00234802106 00234802108 00234802110 00234802112 00234802204 00234802206 00234802208 00234802212 00234802508 00234802510 00234802608

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142579Zimmer Periarticular Plating System
Zimmer, Inc., Oct 22, 2014
K150121Zimmer Periarticular Plating System
Zimmer, Inc., Feb 23, 2015