Recall Z-0314-2020 ยท HRS Plate, Fixation, Bone
Zimmer Biomet, Inc.
Periarticulare Plates - Elbow, Humeral and Ulna Plates, Item Nos. 00234800504 00234800506 00234800508 00234800510 00234800512 00234800514 00234800604 00234800606 00234800608 00234800610 00234800612 00234800614 00234800712 00234800714 00234800716 00234800902 00234800904 00234800906 00234800908 00234800910 00234800912 00234800914 00234801002 00234801004 00234801006 00234801008 00234801010 0023480101
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K150121 | Zimmer Periarticular Plating System Zimmer, Inc., Feb 23, 2015 |