Recall Z-0311-2020 ยท HRS Plate, Fixation, Bone
Zimmer Biomet, Inc.
SM DIS VOL/DORS RAD LAT COL PLATE, Item Nos. 47235801404 47235802102 47235802103 47235802104 47235802106 47235802108 47235802203 47235802204 47235802206 47235802208 47235802210
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K111039 | ZIMMER PERIARTICULAR LOCKING PLATE SYSTEM Zimmer, Inc., May 19, 2011 |