Recall Z-0309-2020 ยท HRS Plate, Fixation, Bone
Zimmer Biomet, Inc.
Prox Lat Tib Lock Plate and Dist Med Tib Lock Plate Item Nos. 47235700306 47235700308 47235700310 47235700312 47235700314 47235700408 47235700410 47235700412 47235700414 47235700416 47235700504 47235700506 47235700508 47235700510 47235700512 47235700514 47235700518 47235700604 47235700606 47235700608 47235700610 47235700612 47235700614 47235700618 47235700706 47235700708 47235700710 47235700714 4
| Date | Oct 10, 2019 |
| Status | Terminated |
| Reason | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
| Root cause | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, |
| Quantity | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| FDA record | Open on fda.gov |