Recall Z-0306-2020 ยท HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer

Zimmer Biomet, Inc.

TM Ankle and Ankle Talus L/R, Item Nos. 00450001100 00450001200 00450001300 00450001400 00450001500 00450001600 00450002100 00450002200 00450002300 00450002400 00450002500 00450002600

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120906ZIMMER TRABECULAR METAL TOTAL ANKLE
Zimmer, Inc., Aug 24, 2012