Recall Z-0305-2020 ยท HRS Plate, Fixation, Bone

Zimmer Biomet, Inc.

ZPLP Fibular Plate, Item Nos. 00234701504 00234701506 00234701508 00234701510 00234701512 00234701516 00234701604 00234701606 00234701608 00234701610 00234701612 00234701614 00234701616 00234701702 00234701704 00234701706 00234701708 00234701710 00234701712 00234701714 00234701716 00234701718 00234701802 00234701804 00234701806 00234701808 00234701810 00234701812 00234701814 00234701816 0023470181

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150121Zimmer Periarticular Plating System
Zimmer, Inc., Feb 23, 2015