Recall Z-0302-2020 ยท KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented

Zimmer Biomet, Inc.

Various Custom Devices, Item Nos. 32855090246 32855090252 32855090256 32855090265 32855090266 32855090270 32855090287 32855090290 32855090291 32855090297 32855090302 32855090317 32855090318 32855090319 32855090327 32855090329 32855090334 32855090336 32855090339 32855090348 32855090351 32855090352 32855090353 32855090354 32855090368 32855090369 32855090375 32855090376 32855090395 32855090396 328550

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.