ClearanceBrief
Recall Z-0300-2020 ยท PKC Prosthesis, Total Anatomic Shoulder, Uncemented Metaphyseal Humeral Stem With No Diaphyseal Incursion, Semi-Constrained

Zimmer Biomet, Inc.

Sidus Stem-Free Shoulder Humeral Head, Item Nos. 0104555380 0104555400 0104555420 0104555440 0104555460 0104555480 0104555485 0104555500 0104555505 0104555520 0104555525

DateOct 10, 2019
StatusTerminated
ReasonNon-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Root causePackaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available,
Quantity5,459,583 total devices
DistributionDistributed nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171858Sidus Stem-Free Shoulder
Zimmer GmbH, Dec 18, 2017