Recall Z-0266-2020 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Datascope Corp.

Low Level Output Cable Interface to Philips Monitor For Use With Cardiosave and CS300 Intra-Aortic Balloon Pumps (IABPs), Part Number 0012-00-1589-03

DateOct 3, 2019
StatusTerminated
ReasonOne lot of "Low Level Output Cable - Interface to Philips Monitor" was received from the supplier with an incorrect cable in the packaging.
Root causeUnder Investigation by firm
ActionOn October 3, 2019, the firm notified customers of the recall via an Urgent Medical Device Recall letter. The letter alerted customers that the incorrect cable may have been supplied. Customers were asked to take the following actions: -Please examine your inventory immediately, remove and quarantine any affected product. - If you have unused affected product, please contact Getinge Customer Service at 1-888-627-8383 (press option 2, then option 1) between the hours of 8 AM and 6 PM Eastern Standard Time to request a material authorization number (RMA) and shipping instructions. Pack the p
Quantity7
DistributionDistributed to CA, GA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.