ClearanceBrief
Recall Z-0219-2020 ยท HSB Rod, Fixation, Intramedullary And Accessories

OrthoPediatrics Corp

Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly referred to as an intramedullary fixation rod (product code HSB). The attachment bolts are one component of the surgical instruments within the system.

DateAug 19, 2019
StatusTerminated
ReasonBinding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.
Root causeNonconforming Material/Component
ActionOn August 19, 2019, the firm notified customers of the issue via emailed notice titled Urgent Medical Device Recall. Customers were informed of the issue of binding between the attachment bolt and the intramedullary nail. The firm asked customers to return all attachment bolts in the field. The issue does not impact intramedullary nails that have been implanted. No additional corrective actions need to be taken for product already implanted. Any questions regarding this recall can be directed to Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at kanderson@orth
Quantity969 total
DistributionWorldwide distributions - US Nationwide distributions including states of AL, AZ, CA, CO, FL, GA, IA, IL, IN, KY, MD, MI, NC, NY, OH, OK, TN, TX, UT, VA, WA, and WI. Country of Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172583OrthoPediatrics PediNail Intramedullary Platform
Orthopediatrics, Inc., May 10, 2018