Recall Z-0257-2020 ยท DQK Computer, Diagnostic, Programmable

GE Healthcare, LLC

CardioLab/ComboLab Recording Systems

DateSep 16, 2019
StatusTerminated
ReasonPotential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.
Root causeDevice Design
ActionGE Healthcare notified customers of the recall on about 09/16/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that certain CLab II Plus Amplifiers used with CardioLab/ComboLab systems have failed the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient. Instructions included that customers could continue to use the system by
Quantity43 (27 US, 16 OUS)
DistributionNationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.