ClearanceBrief
Recall Z-0201-2020 ยท GDW Staple, Implantable

Ethicon Endo-Surgery Inc

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A

DateOct 29, 2019 (initiated Oct 3, 2019)
StatusTerminated
ReasonThe staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.
Root causeNonconforming Material/Component
ActionUrgent Medical Device Recall (Removal) letters were sent to customers beginning 10/3/19. The letters instruct customers to do the following: ACTION REQUIRED: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this voluntary recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to
Quantity255 units
DistributionWorldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160521ECHELON FLEX 60 Powered Plus Compact Articulating Endoscopic Linear Cutter, 280mm
Ethicon Endo-Surgery, LLC, Mar 22, 2016