Recall Z-0265-2020 ยท JAK System, X-Ray, Tomography, Computed
Siemens Medical Solutions USA, Inc
SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
| Date | Sep 24, 2019 |
| Status | Terminated |
| Reason | SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures |
| Root cause | Software design |
| Action | Siemens issued a Customer Safety Advisory Notice to customers via CT072/19 on 9/24/2019. The letter identifies the problem, health risk and action to take: To avoid this issue, please refrain from using the i-Sequence scan mode in combination with the 3x3 mm collimation until this issue has been solved. Please use the i-Sequence with the 3x5 mm collimation instead, as here the correct dose information will be shown. This issue will be resolved with an improved software version syngo CT VA20A_SP2b. The software update will be released, free of charge, with the update CT073/19/S for all affected |
| Quantity | 14 US |
| Distribution | US in the states of: GA, IN, MO, MS, NJ, OH, PA |
| FDA record | Open on fda.gov |