ClearanceBrief
Recall Z-0279-2020 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Prismaflex Control Unit, software versions below 7.21

DateOct 4, 2019
StatusOpen, Classified
ReasonCommunication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Root causeSoftware design
ActionBaxter Healthcare notified customers on about 10/04/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that Prismaflex Control Units with software versions 7.20 and below have the potential to produce communication errors. Instructions included that operators may continue to use the control units until the upgrade is performed, but if a communication error alarm does occur, follow the instructions presented on the graphical user interface and/or in the operators manual. It must be reinforced that it is important to manually return the extracorporeal blood to the
Quantity1,014 units
DistributionNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. International distribution to Australia, Bangladesh, Brunei, China, Ho
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072093PRISMAFLEX SYSTEM, VERSION 3.20
Gambro, Feb 1, 2008