ClearanceBrief
Recall Z-0544-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Gmbh & Co. KG

Accessory kit, Alinity i containing Alinity ci Level Sensor, Bulk Solution

DateSep 16, 2019
StatusOpen, Classified
ReasonPotential reliability issue with the Alinity ci series Level Sensor, Bulk Solution. The development of environmental stress on the bulk solution level sensor can cause cracks.
Root causeDevice Design
ActionAbbott GmbH & Co. KG notified customers on about 09/16/2019, via "Product Correction Urgent Immediate Action Required" letter. Customers were informed that the development of environmental stress on the bulk solution level sensor can cause cracks. Cracks can allow air to enter the fluid line, which may result either a failure to dispense or an incomplete dispense of the bulk solution on Alinity i and Alinity c. Instructions included that if using Alinity i, and you receive one of the following message codes: 1043, 1044, 1072, 1402, 1403. Refer to Appendix B in the letter. If using Alinit
Quantity3,271 units
DistributionNationwide distribution to AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OK, PR, SC, SD, TN, TX, UT, VA, WA, WI. International distribution to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BOTSWANA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.