Recall Z-0207-2020 ยท GEX Powered Laser Surgical Instrument

Candela Corporation

Canister HFC-134a /1000g, ALUM, 15 pack Product Number: 1600-00-0219 Revision .0C Coolant canister for use with surgical laser.

DateOct 24, 2019 (initiated Sep 10, 2019)
StatusTerminated
ReasonThe canister neck bushing may become loose and separate from the body of the canister during use. The issue occurs only when the canisters are used in Candela legacy devices that utilize screw-in or threaded interface type DCD modules
Root causeOther
Action1. Immediately examine your inventory and quarantine the identified lots of the GentleCool Pro Cryogen Canister. 2. Immediately discontinue use and distribution of the identified lots of the GentleCool Pro Cryogen Canister. 3. Please complete and return the enclosed Acknowledgement Form - GentleCool Pro Canister Recall. The completed form must be scanned and emailed to cryogen.usa@candelamedical.com within 24 hours. 4. If you may have further distributed this product, please identify those locations and notify them at once of this product recall. Your communication should include a copy of
Quantity386 cases ( 15 pack canister/ case) (216 additional cases as of 12/09/19)
DistributionStates: NC GA CA MA OH FL IL CO SD NY KS WA MI PA AZ WI TX LA IA MN VA TN SC OR IN WY MD VT ON BC QC NV OUS: Barbados Canada Germany Spain France HR Portugal Ireland Norway Greece NL South Africa Vietnam Hong Kong Japan New Zealand Australia China Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.