Recall Z-0530-2020 ยท FZX Guide, Surgical, Instrument

Zimmer Biomet, Inc.

BALL NOSE GUIDE WIRE 80CM, 100CM. Item Nos. 281001080 281001100 Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.

DateAug 21, 2019
StatusTerminated
ReasonElevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Root causeEnvironmental control
ActionOn September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday throug
Quantity1729
DistributionUS Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.